The Leica SP5 MP is a multiphoton microscope with a DM6000 CFS upright microscope. This microscope is specially useful for intravital deep tissue imaging due to the high tissue penetration depth (several hundred microns) and minimal phototoxicity/photobleaching. An integrated resonant scanner enables image acquisition at video rate (25 fps at 512 x 512 pixels).

Reservation System: Leica TCS SP5 Multiphoton

Location:Biocity, -1floor

20x Leica HCX APO L
• Numerical Aperture: 1.0
• Working Distance: 1.95mm
• Immersion:Water

25x Leica HCX IR APO L
• Numerical Aperture: 0.95
• Working Distance: 2.5mm
• Immersion:Water

Coherent Chameleon Ultra II
  • Tuning range 690 – 1040 nm
  • Coherent Compact OPO extends Ultra II range to 1600nm
Chameleon Vision II
  • Tuning range 690 – 1040 nm

One main beam splitter is usable at a time.


LP505nm Main Beam Splitter
  • SHG440 (440/20nm)
  • CFP (483/32nm)
  • YFP (535/20nm)
  • DsRed (585/40nm)
LP560nm Main Beam Splitter
  • DAPI (457/50nm)
  • FITC (525/52nm)
  • TRITC (585/40nm)
  • Alexa 633 (650/50nm)
  • Photomulitiplier tubes
Leica LAS